Narrow Beam, Stable Wavelength: Specifying Through-Hole UV LEDs for Analyzers and Detectors

News content

A fluorescence detector fails its sensitivity target. The team's instinct is to raise the LED current — but the real problem is beam spill: light is flooding the cuvette holder instead of the sample window, raising background and swamping the signal. The fix is not a brighter emitter. It is a narrower beam angle, and that is a packaging decision made at specification time.

1. The Application Map

These parts are specified into instruments, not lighting fixtures. Tianhui's through-hole UVA series lists anti-counterfeiting detection, money counting and detection, medical analysis, fluorescence reaction, chemiluminescence analyzers, biochemical analyzers, UV photocatalysis and visual light source use. The UVC TO18H series addresses a different bench: water quality inspection, air detection, medical inspection, biochemical analysis and jewellery inspection.

2. Beam Angle Is the Specification That Matters

Two packages can share the same wavelength and the same drive current and still behave nothing alike:

Package Viewing angle (2θ½) Typical fit
F5 (TH-UVxxxJxxx-5) 10° General-purpose instrument optics, moderate collimation
TO46H (TH-UVxx5Jxxx-TO46H) 5.5° Long-path or low-stray-light detection needing tight collimation
TO18H-2P (UVC) 10° Compact UVC sensing heads and detector modules

A 5.5° emitter puts nearly all its flux inside a small cone. In a fluorescence or absorbance path that reduction in stray light often improves the measurement more than doubling the drive current would — and it does so without adding heat to the sample.

3. One Part Number, Seven Wavelengths

The UVA through-hole series is offered at 365, 375, 385, 395, 405, 415 and 425nm, all at a 20mA test current:

  • Forward voltage: 3.0–4.0V (typ. 3.5V)
  • Radiant flux: 4–25mW depending on wavelength and package
  • Spectral half width: 13nm — narrow enough to sit inside an absorption or excitation band
  • Wavelength tolerance: ±3nm
  • Absolute maximums: 180mW dissipation, reverse voltage 5V, operation −20°C to +60°C, soldering 260°C for ≤3s

For UVC work, the TH-UV275WPBY-TO18H-2P delivers 270–280nm (typ. 275nm) at 20mA, with forward voltage 4.8–6.5V, optical power 0.5–2.5mW (typ. 1.1mW), 10° viewing angle, forward current rated to 40mA and 240mW dissipation.

4. Derating Rules to Design Around

Datasheet curves, not intuition, should set the thermal design:

  • Radiant flux falls with ambient temperature. The series holds about 100% at 30°C and drops to roughly 85% at 60°C.
  • Forward voltage drops as temperature rises — about 3.50V at 25°C versus 3.15V at 60°C on the 365nm part, which affects constant-current driver headroom.
  • Peak wavelength drifts with both current and ambient temperature, typically climbing a few nanometres at the top of the operating range. If your filter passband is tight, budget for the shift.
  • These are ESD Class 1 devices. Handle with grounded equipment, never with bare hands, form leads before soldering, and store unopened bags beyond three months only in a sealed container with desiccant — Tianhui leadframes are gold-plated iron alloy and corrode in aggressive atmospheres.

Conclusion

In instrument design the LED is not a light bulb — it is an optical component with a beam cone, a spectral half width, and a temperature coefficient. Specify the viewing angle from your optical path, the wavelength from your excitation or absorption band, and the thermal budget from the derating curves. Do that, and the sensitivity problem you were about to solve with current disappears.

UV LED metal-can emitter illuminating a cuvette inside a laboratory analyzerFigure 1 — Fluorescence / biochemical analysis: narrow-beam through-hole emitter aimed at a sample cuvette.

Banknote under UV inspection light showing fluorescent security featuresFigure 2 — Anti-counterfeiting and currency detection: UV illumination revealing fluorescent security feature